Instrument Function
R&S
®
SMBV100A
214Operating Manual 1176.8016.02 ─ 17
Figure 5-8: Synchronizing a subsequent instrument using the internal 10 MHz reference signal of
the R&S
SMBV
The internal reference oscillator supplies the reference frequency.
Settings:
Source: "Internal"
External f
ref
= 10 MHz (10 MHz REF OUT)
If you have a clean external reference signal with 10 MHz frequency, you can
directly pass it to the output. The signal quality remains the same.
Figure 5-9: Synchronizing instruments by means of an externally applied reference signal having
10 MHz
Settings:
Source: "External"
External Reference Frequency: "10 MHz"
Set the additionally provided parameters, as for example the synchronization band-
width according to the requirements of the application.
External f
ref
= 5/10 MHz (5/10 MHz REF OUT)
If you have an external reference signal with 5 or 10 MHz, you can directly pass it
to the output. The signal quality remains the same.