Remote Control Commands
R&S
®
SMBV100A
743Operating Manual 1176.8016.02 ─ 17
Parameters:
<Impedance> G50 | G1K
G1K
1 kOhm to ground
G50
50 Ohm to ground
*RST: G1K
Example:
INP:CLOC:IMP G1K
sets the clock inputs to 1 kOhm to ground.
Manual operation: See "Impedance Clock Input" on page 140
[:SOURce<hw>]:INPut:MODext:IMPedance
<Impedance>
Sets the impedance for an externally applied modulation signal.
Parameters:
<Impedance> HIGH | G600
HIGH
> 100 k0hm to ground
G600
600 0hm to ground
*RST: HIGH
Example:
INP:MOD:IMP HIGH
sets > 100 kOhm to ground.
Manual operation: See "Ext. Impedance" on page 253
[:SOURce]:INPut:TRIGger:BBANd:SLOPe <Slope>
Sets the active slope of an externally applied trigger signal at the TRIGGER connec-
tors. The setting is effective for both connectors at the same time.
Parameters:
<Slope> NEGative | POSitive
*RST: POSitive
Example:
INP:TRIG:BBAN:SLOP NEG
activates the falling slope of the external trigger signal at the
TRIGGER connectors.
Manual operation: See "Baseband Trigger Input Slope" on page 141
[:SOURce]:INPut:TRIGger:IMPedance <Impedance>
The command sets the impedance of the trigger and control signal inputs and affects:
●
TRIGGER inputs
●
CLOCK inputs