Remote Control Commands
R&S
®
SMBV100A
747Operating Manual 1176.8016.02 ─ 17
Example:
IQ:IMP:IQR 0.5 dB
sets the imbalance.
Manual operation: See "Gain Imbalance - I/Q Mod" on page 300
[:SOURce<hw>]:IQ:IMPairment:LEAKage:I <I>
Sets the carrier offset for the I-channel of the I/Q modulation.
Parameters:
<I> float
Range: -10 to 10
Increment: 0.01
*RST: 0
Default unit: PCT
Example:
IQ:IMP:LEAK:I 3 PCT
sets the leakage for the I-channel to 3 percent.
Manual operation: See "Offset - I/Q Mod" on page 300
[:SOURce<hw>]:IQ:IMPairment:LEAKage:Q <Q>
Sets the carrier leakage amplitude for the Q-channel of the I/Q modulation.
Parameters:
<Q> float
Range: -10 to 10
Increment: 0.01
*RST: 0
Default unit: PCT
Example:
IQ:IMP:LEAK:Q 3 PCT
sets the leakage for the Q-channel to 3 percent.
Manual operation: See "Offset - I/Q Mod" on page 300
[:SOURce<hw>]:IQ:IMPairment:QUADrature[:ANGLe] <Angle>
Sets the quadrature offset for the I/Q modulation.
Parameters:
<Angle> float
Range: -10 to 10
Increment: 0.01
*RST: 0
Default unit: DEG
Example:
IQ:IMP:QUAD:ANGL -5DEG
sets the quadrature offset to -5 degrees.
Manual operation: See "Quadrature Offset - I/Q Mod" on page 300
SOURce Subsystem