Remote Control Commands
R&S
®
SMBV100A
801Operating Manual 1176.8016.02 ─ 17
Example:
PULM:POL INV
selects inverted polarity.
Manual operation: See " Polarity " on page 261
[:SOURce<hw>]:PULM:SOURce <Source>
Selects the source for the pulse modulation signal.
Parameters:
<Source> INTernal | EXTernal
INTernal
The internally generated rectangular signal is used for the pulse
modulation. The frequency of the internal signal can be set in
the SOURce:LFOutput subsystem.
EXTernal
The signal applied externally via the EXT MOD connector is
used for the pulse modulation.
*RST: INTernal
Example:
PULM:SOUR INT
selects the internal modulation source.
PULM:STAT ON
activates the pulse modulation.
Usage: SCPI confirmed
Options: R&S SMBV-K22 (Pulse Modulator)
Manual operation: See "Source" on page 261
[:SOURce<hw>]:PULM:STATe <State>
Activates the pulse modulation.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
PULM:STAT ON
activates pulse modulation.
Manual operation: See "State" on page 261
[:SOURce<hw>]:PULM:TRIGger:EXTernal:GATE:POLarity <Polarity>
Selects the polarity of the Gate signal.
The signal is supplied via the PULSE EXT connector.
Parameters:
<Polarity> NORMal | INVerted
*RST: NORMal
SOURce Subsystem