Remote Control Commands
R&S
®
SMBV100A
805Operating Manual 1176.8016.02 ─ 17
Parameters:
<SBandwidth> WIDE | NARRow
NARRow
The synchronization bandwidth is approx. 50 Hz.
WIDE
The synchronization bandwidth is approx. 350 Hz.
*RST:  n.a. (factory preset)
Example: 
ROSC:SOUR EXT
Selects the external source.
ROSC:EXT:FREQ 10 MHz
Informs the instrument that the external reference has a fre-
quency of 10 MHz.
ROSC:EXT:SBAN WID
Selects wideband setting for synchronization bandwidth.
Manual operation:  See "Synchronization Bandwidth" on page 216
[:SOURce]:ROSCillator[:INTernal]:ADJust:VALue <Value>
Specifies the frequency correction value (adjustment value).
Parameters:
<Value> integer
Range:  0  to  maximum value (see data sheet)
Increment:  see data sheet
*RST:  ---
Example: 
ROSC:ADJ:VAL 456
Sets the adjustment value to 456.
Manual operation:  See "Adjustment DAC Value" on page 217
[:SOURce]:ROSCillator[:INTernal]:ADJust[:STATe] <State>
Determines whether the calibrated (OFF) or a user-defined (ON) adjustment value is
used for fine adjustment of the frequency.
If user-defined values are used, the instrument is no longer in the calibrated state.
However, the calibration value is not changed and the instrument resumes the calibra-
ted state after sending the command :SOURce:ROSCillator:INTernal:ADJust:
STATe OFF.
Parameters:
<State> 0 | 1 | OFF | ON
*RST:  n.a. (factory preset: 0)
SOURce Subsystem