Remote Control Commands
R&S
®
SMBV100A
820Operating Manual 1176.8016.02 ─ 17
If a bit in the ENABle part is 1, and the correesponding EVENt bit is true, a positive
transition occurs in the summary bit. This transition is reportet to the next higher level.
Parameters:
<Enable> string
Example:
STAT:QUES:ENAB 1
Problems when performing an adjustment cause an entry to be
made in the sum bit.
:STATus:QUEStionable[:EVENt] <Event>
Queries the content of the EVENt part of the STATus:QUEStionable register. This
part contains information on the actions performed in the instrument since the last
readout. The content of the EVENt part is deleted after being read out.
Parameters:
<Event> string
Example:
STAT:QUES:EVEN?
queries the Status:Questionable:Event register.
:STATus:QUEStionable:NTRansition <Ntransition>
Sets the bits of the NTRansition part of the STATus:QUEStionable register. If a bit is
set, a transition from 1 to 0 in the condition part causes an entry to be made in the
EVENt part of the register.
Parameters:
<Ntransition> string
Example:
STAT:QUES:NTR 0
a transition from 1 to 0 in the condition part of the STA-
Tus:QUEStionable register does not cause an entry to be made
in the EVENt part
:STATus:QUEStionable:PTRansition <PTransition>
Sets the bits of the NTRansition part of the STATus:QUEStionable register. If a bit is
set, a transition from 1 to 0 in the condition part causes an entry to be made in the
EVENt part of the register.
Parameters:
<PTransition> string
Example:
STAT:QUES:PTR 32767
all transitions from 0 to 1 in the condition part of the STA-
Tus:QUEStionable register cause an entry to be made in the
EVENt part
STATus Subsystem