Remote Control Commands
R&S
®
SMBV100A
836Operating Manual 1176.8016.02 ─ 17
:SYSTem:PRESet
Triggers an instrument reset. It has the same effect as:
●
The PRESET key
●
The *RST command
For an overview of the settings affected by the preset function, see Chapter 5.2.2,
"Default Instrument Settings - Preset Key", on page 118.
Example:
SYST:PRES
All instrument settings (also the settings that are not currently
active) are reset to their default values.
Usage: Setting only
:SYSTem:PROTect<ch>[:STATe] <State>[, <Key>]
Activates and deactivates the specified protection level.
Suffix:
<ch>
.
Indicates the protection level.
See also Chapter 5.2.3.17, "Protection", on page 146.
Parameters:
<State> select
*RST: n.a. (factory preset: 1)
Setting parameters:
<Key> integer
The respective functions are disabled when the protection level
is activated. No password is required for activation of a level. A
password must be entered to deactivate the protection level.
The password for the first level is 123456.
Example:
// to activate protection level
SYSTem:PROTect1:STATe 1
// internal adjustments or hostname cannot be changed
// to unlock protection level 1
SYSTem:PROTect1:STATe 0,123456
// internal adjustments are accessible
Manual operation: See "Protection Level/Password" on page 147
:SYSTem:REBoot
Restarts the firmware and the operating system.
Usage:
Event
:SYSTem:RESTart
Restarts the firmware. The operating system remains active.
SYSTem Subsystem