Sirona Dental Systems GmbH 4 Operation
Operating Instructions SINIUS / SINIUS CS 4.5 Dentist element
båÖäáëÜ
4.5.12 SPRAYVIT M multifunctional syringe
SPRAYVIT M multifuncti onal syringe
The multifunctional syringe is used to clean the treatment areas and blow
them dry. It supplies air and preheated water media. The heating
cartridge for the water is located in the handpiece.
Multifunctional syringe, observe the Opera ting Instructions
4.5.12.1
Safety instructions
Safety instructio ns
The SPRAYVIT M is equipped with extensive safety monitoring functions.
However, please observe the following information.
Multifunctional syringe, changing hoses
Multifunctional syringe, switch-on de lay
Multifunctional syringe, flow rate
Multifunctional syringe, air outlet
IMPORTANT
Also observe the operating instructions of the SPRAYVIT M.
CAUTION
After changing the SPRAYVIT M hose, no cooling water for the
SPRAYVIT M heating cartridge flows until the hose is completely filled.
The patient may be scalded by the emission of hot steam. The heating
cartridge can overheat and be destroyed.
➢ After changing the hose of the SPRAYVIT M multifunctional syringe,
press the
Water
key briefly and repeat until there is an ample supply
of water in the hose before treating the patient.
IMPORTANT
Heating cartridge switch-on delay
To minimize the risk of scalding, the SPRAYVIT M water heater is not
activated for several seconds after the treatment center is switched on
and after a hose change when the
Water
key is initially pressed.
CAUTION
If the flow rate is insufficient, hot water may be emitted by the
SPRAYVIT M.
The patient could thus be scalded.
➢ Check the water flow rate prior to use.
➢ Check the flow rate at least once a month and whenever you
suspect that it may be insufficient as described in the section
"Checking the flow rate of the SPRAYVIT multifunctional
syringe" [ → 189]. Clean the nozzle according to the SPRAYVIT M
Operating Instructions.
IMPORTANT
Air outlet during hose change
If the SPRAYVIT M is removed from the instrument hose while the
treatment center is in use, air will escape from the hose coupling. You
should therefore switch off the treatment center before changing a hose.