Sirona Dental Systems GmbH 7 Malfunctions
Operating Instructions SINIUS / SINIUS CS 7.1 Error messages
båÖäáëÜ
7
Malfunctions
Malfunct ions
7.1
Error messages
Error messages
Any error states of the treatment center that cannot be immediately
recognized by the user and must nevertheless be corrected are displayed
on the touchscreen. If an error state exists, this is indicated by a warning
triangle (A) in the status line of the touchscreen.
Touch the warning triangle (A) in the footer of the touchscreen.
The error code(s) is(/are) displayed.
The error codes have the following meaning:
Error table, without HF surgery, without hydro colloid, without safety switch for chair frame
Code Error/fault Description Action
10 Flow rate of SPRAYVIT M
on dentist element too low
The water flow rate is below the limit
value. The heater of the SPRAYVIT M
is no longer heated to prevent
scalding by overheated water.
Clean the outlet nozzle of the
SPRAYVIT, see the "SPRAYVIT M
Operating Instructions" and perform
the flow test, see "Checking the flow
rate on the SPRAYVIT M
multifunctional syringe".
11 Flow rate of SPRAYVIT M
on assistant element too
low
12 Tumbler temperature fuse
defective
The temperature fuse was blown by
overheating. The water in the tumbler
filling unit is no longer heated.
Inform your service engineer.
13 Battery of wireless foot
control empty
Operation of the treatment center no
longer possible due to fully depleted
battery.
See "Changing the battery of the
wireless foot control" [ 215].
14 Disinfectant consumption
too low
The disinfectant tank was not emptied
since its last filling even though many
disinfection cycles were performed.
Inform your service engineer.
15 Water feed too low The maximum filling time for the
mixing tank is exceeded.
Change the water filter, see
"Changing the water and air
filters" [ 201].
Check the water pressure of the
drinking water network.
17 Emergency pump operation Malfunctioning of pump sensors in the
water unit.
Inform your service engineer.
18 Chair movement outside of
the permissible movement
limits or chair drive position
not plausible
If the chair is located outside of the
permissible movement limits or the
position is not plausible, the chair
stops moving.
Inform your service engineer.
19 Injection valve for
disinfectant defective
No disinfectant is added if the
injection valve is defective.
Inform your service engineer.