2 Safety instructions Sirona Dental Systems GmbH
2.4 Media quality Operating Instructions SINIUS / SINIUS CS
2.4
Media quality
Media quality
The air and water supplies must meet the requirements specified in the
installation instructions. Use only drinking water and dry, oil-free and
hygienically clean air for the water and air supplies of the treatment
center.
Disinfection system, optional
To ensure compliance with the medical and national legal requirements
for water from treatment centers, Sirona recommends equipping the
treatment center with a disinfection system. If you decide to operate the
treatment center without a disinfection system, you must make alternative
arrangements to ensure good water quality.
As the owner of the treatment center, you are generally responsible for
the water quality.
For this reason, you should check the water quality at regular intervals,
see "Microbiological inspection of the water" [ 170]. Please contact your
specialized dealer or your relevant dental association for the respective
national requirements and measures. Where necessary, you must make
alternative arrangements to ensure good water quality if you decide to
operate the treatment center without a disinfection system.
Immunosuppressed patients, with sterile solution
Highly immunosuppressed patients should not come in contact with water
from the treatment center. The use of sterile solutions is recommended.
2.5
Connection to the public drinking water system
Connecti on to the publ ic drinking water s ystem
Free discharge where no disinfection system exists
Provided it is equipped with a disinfection system, the treatment center
provides free discharge in the water supply (isolating distance 20 mm).
This isolates the public drinking water system from the water supplies
following free discharge. The requirements of EN 1717 are thereby
fulfilled.
The treatment center meets the requirements of the DVGW (German Gas
and Water Association).
It is intrinsically safe in accordance with worksheet W540 and therefore
fulfills the requirements of W270 and KTW (plastics in waterways). The
German Gas and Water Association (DVGW) approval can be viewed on
a label next to the rating plate.
Structural measures if no disinfection system is available
Provided the treatment center is not equipped with a disinfection system,
structural measures may be required beyond the unit to protect the public
drinking water system.
Please always adhere to the national requirements with regard to
connecting treatment centers to the public drinking water system.
2.6
Maintenance and repair
Maintenance and repair
Authorized technical personnel and
spare parts
As manufacturers of dental medical units and in the interest of the
operational safety of your system, we stress the importance of having
maintenance and repair of this unit performed only by ourselves or by
agencies expressly authorized by us. Furthermore, failed components
must always be replaced with original spare parts.
Certificate, with modification n ote
We suggest that you request a certificate showing the nature and extent
of the work performed from those who carry out such work; it must contain