Sirona Dental Systems GmbH 4 Operation
Operating Instructions SINIUS / SINIUS CS 4.1 Starting up the treatment center
båÖäáëÜ
4
Operation
Operation
4.1
Starting up the treatment center
Starting up the treatment center
4.1.1 Initial startup
Initial startup
Sanitation must be performed prior to initial startup of your treatment
center.
This is done by filling all water-bearing lines with DENTOSEPT P
concentrated disinfectant to reduce the exposure of the water to bacteria.
SINIUS reference
If the service engineer skipped the sanitation procedure after installing
your treatment center based on an agreement with you or sanitation has
not been performed for more than one week, please perform sanitation
yourself. Refer to "Sanitizing the treatment center" [ → 207] for more
information.
Sanitation takes approx. 24 hours.
4.1.2 Switching the treatment center on/off
Switching the treatment center on/o ff
The treatment center is equipped with a standby system for enhanced
convenience when switching it on and off at the dentist element.
The treatment center thus features a power switch at the base of the chair
and a main switch on the dentist element.
4.1.2.1
Power switch
Power switch
The power switch connects the treatment center to the power supply.
During longer periods of disuse, the treatment center should be
disconnected from the power supply. It then no longer consumes any
energy.
The power switch contains an automatic device fuse.
Connecting the treatment center to the power supply
✔ The treatment center is installed by authorized technical personnel
according to the "Installation Instructions".
➢ Turn on power switch A.
The treatment center is connected to the power supply.
Disconnecting the treatment center from the power supply
✔ The treatment center is shut down, see "Switching the treatment
center off" (below).
➢ Turn power switch A off.
The treatment center is disconnected from the power supply.