Sirona Dental Systems GmbH 2 Safety instructions
Operating Instructions SINIUS / SINIUS CS 2.7 Trouble-free operation
båÖäáëÜ
any changes in rated parameters or working ranges (if applicable), as well
as the date, the name of the company and a signature.
Maintenance intervals Despite the outstanding quality of your treatment center and regular care
by the practice team, in the interest of operational safety, it is essential to
have preventive maintenance performed at scheduled intervals.
In order to ensure the operational safety and reliability of your treatment
center and to avoid damage due to natural wear, you as the system owner
must have your system checked regularly by an authorized service
engineer from your dental depot. Furthermore, safety checks must be
performed. Please contact your dental depot to obtain a maintenance
offer. For more information, please refer to "Maintenance by the service
engineer" [ → 217].
2.7
Trouble-free operation
Trouble-free operation
Use of this unit is permissible only if it works properly without
malfunctions. If trouble-free operation cannot be ensured, the unit must
be taken out of service, checked by authorized technicians for
malfunctions and, if necessary, repaired.
2.8
Vacuum system
Vacuum system
Suction of aluminum or other metal oxides prohibited
The suction removal of aluminum and other metal oxides from blasting
devices via the amalgam separator and the automatic separator installed
in the treatment center is prohibited! This would result in extreme wear
and clogging of the vacuum and drain lines.
Separate suction
A separate vacuum system must be used in connection with metal oxide
blasting devices. Treatment centers equipped with a central wet suction
system are generally suitable for suction removal of the above material.
However, make sure to observe the instructions provided by the
manufacturer of your vacuum system.
Suction of steel abrasives permissible
No restrictions apply when using salt blasting devices in connection with
Sirona treatment centers. However, in such cases, make sure that the
system is subsequently flushed with an adequate amount of water.
2.9
Patient chair
Patient chair
Please observe the maximum load capacity of 165 kg for the patient
chair.
The weight distribution complies with ISO 6875. The mechanical stability
test is performed with a multiple safety factor according to IEC 60601-1.
The maximum permissible weight of accessories mounted on the patient
chair is 5 kg.
The patient’s arms and legs must be resting on the upholstery of the
chair.