63 22 726 D3561
D3561.201.01.03.02 05.2012
161
Sirona Dental Systems GmbH 4 Operation
Operating Instructions SINIUS / SINIUS CS 4.13 Configuration of the treatment center (setup)
båÖäáëÜ
➢ Touch the
Start program operating mode
key.
If the key is highlighted orange, the start program with instrument
functions is displayed.
Changing setup program page
Changing setup program page
➢ Change to the next setup program page.
4.13.4.4
Adjust cleaning agent mixture for chemical suction hose cleaning
Adjust cleaning agent mixture for chemical suction hose cleaning
The vacuum system can be cleaned by pumping water into a tank behind
the receptacle and extracting it from there. A cleaning agent is added to
the water if the dental treatment center is equipped with the chemical
suction hose cleaning option. For more information, please refer to the
section “Carrying out suction hose cleaning” [ → 192]..
It is possible to set how much cleaning agent should be added to the
water for chemical suction hose cleaning. The quantity is dependent on
the cleaning agent used and the type of treatment. Please observe the
manufacturer’s instructions for the cleaning agent.
➢ Use the – and + keys to adjust the cleaning agent mixture for
chemical suction hose cleaning (0 to 5%).
The adjusted percentage is displayed on the touchscreen.
4.13.4.5
Key for showing/hiding white screen on SIVISION monitor
Key for sh owing/hiding white screen on SIVISION moni tor
If the treatment center has no X-ray viewer, however, is equipped with a
SIVISION monitor, the SIVISION monitor can be set to the white screen
mode.
➢ Touch the
White screen
key.
If the key is highlighted orange, the
White screen
key is displayed
in the
Start program
.
4.13.4.6
Linking the spray aspirator to the 4-way foot switch
Linking the spray aspirator to the 4-way foot switch
Saliva ejector function
A setting can be made to enable interruption and/or reactivation of the
suction flow of the spray aspirator by pressing the 4-way foot switch on
the base of the chair in any direction. This function cannot be used on the
saliva ejector. Also observe the safety information, see "Suction
handpieces".
If you deposit the spray aspirator in its holder while the suction flow is
interrupted, the suction flow is automatically restarted when you pick it up
again.