4.5 Dentist element Operating Instructions SINIUS / SINIUS CS
with the motor stop function activated. The three beeps that are issued
following activation of the auto-reverse function continue to sound.
4.5.9.4
Performing manual measurements with the file clamp
Performing manual measurements with the file clamp
For endodontic examination, a manual measurement can be performed
using the file clamp and a root canal file.
✔ The treatment center is prepared for manual measurements using
the file clamp, see section "Preparing to use the ApexLocator" [ → 94].
✔ The
Start
sub-screen is displayed on the touchscreen.
1. Touch the
Manual Measurement with File Clamp
key.
If this function is activated, the
Manual measurement with file
clamp
key is highlighted in orange.
The distance display is shown on the touchscreen.
2. If the distance to the apex is to be indicated with an acoustic signal,
press the
Apex distance signal tones
button.
If the button is colored orange, signal tones will be issued as well
as the graphical distance display. The intervals between the
signal tones vary according to the measured distance to the
apex.
3. Attach a root canal file to the file clamp.
4. Short the electrical measurement system. Hold the clamped files
directly against the mucosal electrode. This cancels out any
inaccuracies caused by skips in impedance in the measurement
setup (standardization).
If a short signal is heard and the distance display appears with no
bars, the standardization was successful. If not, check the
electrical connections for signs of damage.
5. Place the mucosal electrode in the patient's mouth and perform the
measurement.
The measured root canal depth is displayed by a colored bar in
the distance display. For more information, please refer to
"Distance display" [ → 96].
IMPORTANT
Preventing incorrect measurements
When performing the apex measurement, wear non-conductive gloves
to avoid measurement errors caused by unwanted leakage current.
During the measurement, the root canal file must not come into contact
with the patient's mucosa or the mucosal electrode. It is recommended
to perform the measurement using a cofferdam.