Sirona Dental Systems GmbH 5 Care, cleaning and maintenance by the practice team
Operating Instructions SINIUS / SINIUS CS 5.4 Vacuum system
båÖäáëÜ
5.4.3 Sterilize and disinfect the suction handpieces
Sterilize and disinfect the suction handpieces
Sterilization/disinfection
All parts of the suction handpieces can be sterilized and thermally
disinfected.
Greasing suction handpieces
The disconnection points of the suction handpieces must be greased
following each thermal disinfection or sterilization process and once a
week.
Grease
Without third suction hose option
1. Pull the handpiece of the spray aspirator and the saliva ejector off the
suction hoses.
2. Take apart the handpiece of the spray aspirator at its joints.
3. Grease the disconnection points and O rings of the handpieces.
5.4.4 Cleaning and disinfecting the suction hoses
Cleaning and disinfecting the suction hoses
Without third suction hose option
The hoses of the spray aspirator and the saliva ejector as well as the
connection hose to the water unit can be pulled off for rinsing under
running water.
Filter inserts are inserted between the suction hoses and the assistant
element to filter out solid particles. Depending on the treatment involved,
it may also periodically be necessary to remove the trapped solid particles
(e.g. amalgam) from the collector due to a decrease in the suction power.
E
D
C
B
A
134°C 273,2°F
3 min
A Suction cannula
B Suction handpiece
C Handpiece of saliva ejector
D Surgical suction cannula
E Intermediate piece
CAUTION
Unsuitable greases
Greases that are not food safe may endanger the patient's health
Rubber materials such as e.g. O-rings are corroded by unsuitable
greases.
Do not under any circumstances use vaseline or silicone greases.
Use only greases approved by Sirona.