Sirona Dental Systems GmbH 5 Care, cleaning and maintenance by the practice team
Operating Instructions SINIUS / SINIUS CS 5.5 Components of the water unit
båÖäáëÜ
6. Place the container in the special packaging and ship it for disposal
or authorize a certified waste management company.
Installing the amalgam rotor
Use only original Sirona accessories. Never use a used or recycled
amalgam rotor.
To reorder the amalgam rotor, see "Spare parts and
consumables" [ → 221].
1. Grease the O-ring (E) on the lower part of the amalgam separator.
For greases, see "Care and cleaning agents" [ → 170].
2. Insert the new amalgam rotor (C) in the lower part of the amalgam
separator (B).
3. Hold the lower part of the amalgam separator (B) so that the latching
noses of the bayonet catch are positioned transverse to the water
unit. Screw the lower part of the amalgam separator (B) onto its upper
part by rotating it clockwise.
4. Close the flap (A).
5. In Germany: Document replacement of the amalgam rotor in the
"Amalgam separator operator's log D3181 II"
Worldwide: Write documentation according to the relevant national
regulations.
Amalgam separator operator's log
In Germany, users are obligated by law to keep a system log for the
amalgam separator. This log is included with the treatment center. Please
note the user duties as described in the operations log:
● Document the replacement of the amalgam rotor
● Check the function of the amalgam separator system annually
● Arrange for a 5-year inspection
5.5.7 Emptying the sediment container
Emptying the sediment container
In addition to other solid particles, a large proportion of the amalgam
residues are trapped in the sediment container by gravitational force.
Empty the sediment container in cycles that are appropriate for your work
method, but at least every 4 weeks.
The sediment container is installed only if neither an amalgam separator
nor a wet suction device is installed.
NOTICE
Amalgam separator message
If the message
Amalg
is still displayed on the touchscreen and the
acoustic signal persists after the amalgam rotor has been inserted, the
lower part of the amalgam separator is not properly locked.