Sirona Dental Systems GmbH 5 Care, cleaning and maintenance by the practice team
Operating Instructions SINIUS / SINIUS CS 5.3 Instruments and instrument hoses
båÖäáëÜ
5.3
Instruments and instrument hoses
Instruments and instrument hoses
5.3.1 Rinsing water lines
Rinsing water lines
Microorganisms can grow in the water lines of the treatment center. Use
a large amount of water for rinsing the lines prior to starting patient
appointments.
➢ Activate the cuspidor flushing of the cuspidor bowl for at least one
minute.
5.3.2 Purge water lines (purge function)
Purge water lines (purge function)
To reduce the amount of germs, the water lines of the water-carrying
instruments of the dentist element and the SPRAYVIT M multifunctional
syringe of the assistant element are purged with water.
For the purge function, individual instruments are removed and held over
the cuspidor for purging. The water lines of all removed instruments are
then purged simultaneously. Press the water key on the SPRAYVIT M
unit for purging.
It is also possible to purge the water lines automatically, see "Purging
water lines automatically (Autopurge function)" [ → 182].
Preparation
The following preparations should be made before you begin to purge the
water lines.
1. Set all of the instruments to be purged to their maximum water flow
rate.
2. Put all instruments in place.
Opening the Purge screen
✔ The
Start program
is displayed on the touchscreen.
1. In the
Start program without instrument functions
: Touch the
Sub-
screen
key.
In the
Start program with instrument functions
: Press the
Chair
program change key.
The
Start
sub-screen is displayed.