5 Care, cleaning and maintenance by the practice team Sirona Dental Systems GmbH
5.3 Instruments and instrument hoses Operating Instructions SINIUS / SINIUS CS
The silicon sleeve is the exception here. It must be changed after each
patient. The silicon sleeve must be sterilized prior to initial use.
To reorder the silicon sleeves, see "Spare parts and
consumables" [ 221].
Sterilize the root canal files according to the manufacturer's instructions.
5.3.4.3
Disinfecting/sterilizing Mini L.E.D. curing light
Disinfecting/sterilizing Mini L.E.D. curing light
Care and cleaning agents
1. Removing the connection cable of the Mini L.E.D. by turning the
handpiece.
2. Pull out the light guide and remove the glare shield.
3. Sterilize the light guide and the glare shield at 134° C, 2 bar for 3 min.
4. Disinfect the handpiece of the Mini L.E.D.
5. Screw the sterilized light guide and glare shield back onto the Mini
L.E.D.
6. Reconnect the handpiece of the Mini L.E.D. to the connecting cable.
The following points should also be observed when operating the Mini
L.E.D.:
Always use the glare shield to protect your eyes.
Check the light guide after each use. Make sure that the light guide is
in perfect condition.
No composite material residues may be present on the light guide.
Immediately remove any residue.
If you find any damage, replace the light guide, since damage will
impair its performance considerably.
5.3.4.4
Cleaning/disinfecting the SIROCAM intraoral camera
Cleaning /disinfecting the S IROCAM intraoral camera
Due to its shape, the SiroCam AF intraoral camera takes hygienic
requirements into account and therefore has no spots that are difficult to
reach. It can be wipe disinfected.
Care and cleaning agents
NOTICE
Approved care and cleaning agents
Use only care and cleaning agents which have been approved by Sirona
(see "Care and cleaning agents" [ 170]).
134°C 273,2°F
3 min
SiroCam AF
NOTICE
Approved care and cleaning agents
Use only care and cleaning agents which have been approved by Sirona
(see "Care and cleaning agents" [ 170]).