Sirona Dental Systems GmbH 5 Care, cleaning and maintenance by the practice team
Operating Instructions SINIUS / SINIUS CS 5.4 Vacuum system
båÖäáëÜ
5.4.5 Thermodisinfecting suction hoses
Thermodisinfecting suction hoses
The hoses the treatment center is equipped with as a standard feature
cannot be thermally disinfected. Thermally disinfectable suction hoses
are available as special accessories. They are marked with a turquoise
ring, see item A in the drawing below.
Before thermally disinfecting the suction hoses, you must remove them
from the assistant element and clean the filter inserts; see "Cleaning the
suction hoses" [ 195].
The suction hoses are removed and cleaned.
1. Pull the handpiece (B) off of the spray aspirator hose.
2. Grease the O-rings (C). For greases, see "Care and cleaning
agents" [ 170].
3. Plug the ends of the handpiece suction hoses onto the hose holders.
Use the adapter (D) for the saliva ejector hose.
4. Plug the hose holders onto the rails in the (Miele) thermodisinfector
and place the suction hoses on the wire basket provided for that
purpose.
5. Thermally disinfect the suction hoses, filter housings and collectors at
max. 93°C.
To reorder the hose holders for the Miele thermodisinfector, see "Spare
parts and consumables" [ 221].
93°C 197°F