EasyManua.ls Logo

Sirona Sinius - Apexlocator; Care For, Disinfect and Sterilize the Treatment Instruments; Cleaning and Disinfecting;Sterilizing the Components of the; Treatment Instruments

Sirona Sinius
248 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
63 22 726 D3561
D3561.201.01.03.02 05.2012
187
Sirona Dental Systems GmbH 5 Care, cleaning and maintenance by the practice team
Operating Instructions SINIUS / SINIUS CS 5.3 Instruments and instrument hoses
båÖäáëÜ
5.3.4 Care for, disinfect and sterilize the treatment instruments
Care for, disi nfect and sterilize the treatm ent instruments
5.3.4.1
Treatment instruments
Treatment instruments
The procedures required for the following treatment instruments are
described in the following separate operating instructions:
List of treatment instruments, without SL motor and implant and HF surgery
●BL motor
BL ISO S/C motor
Straight and contra-angle handpieces in various versions
High-speed handpieces
SPRAYVIT M multifunctional syringe
SIROSONIC TL scaler
5.3.4.2
Cleaning and disinfecting/sterilizing the components of the
ApexLocator
Cleaning and disinfecting/sterilizing the components of the ApexLocator
Care and cleaning agents
Disconnect the mucosal electrode from the connection cable. The metal
hook can be sterilized, and the connection cable can be disinfected by
wiping.
The file clamp for manual measurement can be sterilized with the
connection cable.
The apex adapter and its connection cable can be disinfected by wiping.
ApexLocator, no disinfectant between contacts
NOTICE
Approved care and cleaning agents
Use only care and cleaning agents which have been approved by Sirona
(see "Care and cleaning agents" [ 170]).
134°C 273,2°F
3 min
134°C 273,2°F
3 min
IMPORTANT
To guarantee conductivity, no disinfectant must be allowed to penetrate
the electrical contacts.

Table of Contents

Other manuals for Sirona Sinius

Related product manuals