4 Operation Sirona Dental Systems GmbH
4.4 Patient chair Operating Instructions SINIUS / SINIUS CS
4.4
Patient chair
Patient chair
4.4.1 Safety instructions
Safety instructions
Patient chair, chair movements
Maximum load capacity o f patient chair
Patient chair, movement range
Patient chair, interlocked when instrument is active
If chair movement is permanently blocked, please contact your service
engineer.
CAUTION
The free space under the patient couch and up to the water unit can be
decreased due to chair movements.
Parts of the patient's or user's body may be pinched or crushed.
Do not allow any limbs to be caught between the chair upholstery
and the chair base. Please make sure that the patient’s arms and
legs rest on the upholstery of the chair.
Do not place any objects on the base of the chair.
CAUTION
The maximum load capacity of the patient chair is 165 kg acc. to ISO
6875 (tested with multiple safety acc. to IEC 60601-1).
If the maximum load capacity is exceeded,a risk of damage to the
treatment chair and injury of the patient exists.
Never allow any persons who weigh more than 160 kg to sit on the
patient chair.
The maximum additional weight of accessories mounted on the
patient chair is 5 kg.
CAUTION
Objects protrude into the movement range of the chair.
There is a risk of crushing the patient and damaging the objects.
Make sure that no objects such as e.g. windows, drawers or other
devices protrude into the movement range of the treatment center.
IMPORTANT
Chair interlock
As long as a treatment instrument is activated, all functions for moving
the patient chair are disabled for safety reasons.