Sirona Dental Systems GmbH 5 Care, cleaning and maintenance by the practice team
Operating Instructions SINIUS / SINIUS CS 5.5 Components of the water unit
båÖäáëÜ
5.5.8 Cleaning the filter insert of the wet suction device
Cleaning the fi lter insert of the wet suct ion device
The automatic separator and amalgam separator or sediment container
are not installed in the water unit for wet suction. The separation of air and
water and amalgam separation are performed centrally in this case.
In order to ensure that larger solid particles nevertheless cannot enter the
separating unit, the vacuum line in the water unit is equipped with a filter.
The filter insert must be cleaned when the suction power decreases.
Wear gloves, risk of contamination
Cleaning the vacuum system
The vacuum system should be cleaned before cleaning the filter insert of
the wet suction device; see "Cleaning suction hoses using adapter in the
cuspidor" [ → 193].
Start program with or without instrument functions
Opening Start sub-screen
✔ The
Start program
is displayed on the touchscreen.
➢ In the
Start program without instrument functions
: Touch the
Sub-
screen
key.
In the
Start program with instrument functions
: Press the
Chair
program change key.
The
Start
sub-screen is displayed.
Extracting residual water
As long as the treatment center is switched on, water mains in the
vacuum line for technical reasons. Therefore, in order to clean the filter
insert this water must be completely extracted beforehand. Otherwise the
remaining water will flow out when the filter housing is opened. When the
treatment center is switched off at its main switch, the remaining water is
automatically extracted.
➢ Touch the
Extract residual water
key.
As long as the key is lit up in orange, water is being extracted
from the water unit. A slurping noise indicates that the water unit
is completely empty. When the key turns gray the extraction
process is finished.
WARNING
Wear gloves when performing the following work.