63 22 726 D3561
24 D3561.201.01.03.02 05.2012
3 Unit description Sirona Dental Systems GmbH
3.2 Technical data Operating Instructions SINIUS / SINIUS CS
3.2
Technical data
Technical data
Model designation, power ratings
Unit classifications, without HF surgery
Operating and transport conditions
Model designation: SINIUS / SINIUS CS
Power connection: 100 - 230 V AC ± 10%
50/60 Hz
Rated current: 2.2 A at 230 V
4.35 A at 115 V
5.0 A at 100 V
also max. 6 A for external devices
Overvoltage category: 2 acc. to IEC 60664-1
Average power consumption
(for dimensioning an air
conditioning system):
0.25 kW
Power consumption in
Standby mode:
3 W (without internal mini PC)
Main building fuse: Type B automatic circuit breaker
100 - 115 V AC: 20 A slow-blow
230 V AC: 16 A slow blow
Protection class: Class I equipment
Device class in accordance
with Directive 93/42/EEC:
Class IIa equipment
Degree of protection against
electrical shock:
Type B applied parts
External intraoral camera SiroCam AF.
Namely:
Application part type BF
Degree of protection against
ingress of water:
Ordinary equipment (without protection
against ingress of water)
The foot control has an IP X1 degree of
protection against liquids (drip-proof).
Mode of operation: Continuous operation with intermittent
loading corresponding to the dental
mode of working. [ 18]
Permanently connected unit.
Transport and storage
conditions:
Temperature: -40 ℃ – +70
(-40.00 ℃ – 70.00 )
Relative humidity: 10% 95%
Barometric pressure: 500 hPa
1060 hPa