4.5 Dentist element Operating Instructions SINIUS / SINIUS CS
1. The instruments must be removed from the instrument holder. To do
this, remove all instruments one after the other and allow these to
overhang facing forward.
2. Lift the instrument holder by the front edge until the clamps are
released and the holder can be removed.
3. Set the spray amount by turning the screw.
4. Check the set spray amount with a burr drive. Correct the setting if
necessary.
5. First insert the rear edge of the instrument holder into the groove on
the dentist element. Then push the holder forward and down until it
clicks into place.
6. Place the instruments in their holders. Make sure that the instrument
hoses are located in the guide rolls of the swivel arm.
4.5.8.7
Preparing for use of NaCl saline solution
Preparing for use of NaCl saline solution
The peristaltic pump prepares a sterile saline solution instead of spray
water for cooling.
The peristaltic pump hose is a disposable item. To reorder it, see "Spare
parts and consumables" [ 221].
The drive of the peristaltic pump and a receptacle for the NaCl bottle
holder are attached to the dentist element. Please contact your local
distributor if necessary.
A new peristaltic pump hose is available.