Sirona Dental Systems GmbH 5 Care, cleaning and maintenance by the practice team
Operating Instructions SINIUS / SINIUS CS 5.1 Basics
båÖäáëÜ
4. Immerse the tester in the filled container for 30 seconds.
The cardboard disk with the nutrient medium will now absorb 1 ml
of the water sample.
5. Remove the tester from the container. Shake out any excess water.
Empty the tank.
6. Place the tester in the container for incubation either for two days at
a room temperature of 20°C/68°F or for 24 hours at a temperature of
35°C/95°F.
7. Count all germs found on the surface of the tester.
Link to sanitizing the treatment center, SINIUS/SINIUS CS
If the number of germs significantly exceeds 100, then sanitation is
required; see "Sanitizing the treatment center" [ 207].
30 s
48 h – 20 °C / 68 °F
24 h – 35 °C / 95 °F