Sirona Dental Systems GmbH 4 Operation
Operating Instructions SINIUS / SINIUS CS 4.5 Dentist element
båÖäáëÜ
Dentist element, hoses, entry/exit
SINIUS CS dentist element
In the swivel arm version, the dentist element is attached to the water unit
with a flexible support arm. The support arm is held in place by a
pneumatic locking brake. A capacitive key for releasing the brake is
integrated in both handles. The key responds when the handles are
touched.
1. Grip a handle with your hand.
This activates the capacitive key. The emergency brake is
released with the sound of pressurized air.
2. Move the dentist element to the required position and release the
handle.
The dentist element is now positioned securely.
Please note that you must wait for at least two seconds before releasing
the brake in the support arm again.
Swivel arm dentist element, collision, manual movement
WARNING
The patient's feet may get caught in the instrument hoses of the dentist
element when he enters or leaves the patient chair.
The patient may trip or fall.
➢ Turn the dentist element outward before the patient enters or leaves
it.
WARNING
The SINIUS CS dentist element could be positioned within the
movement range of the patient chair.
When the patient chair is moved, the patient may collide with the SINIUS
CS dentist element or its support arm. The patient is at risk of being
crushed.
➢ Before moving the patient chair, position the SINIUS CS dentist
element in such a way that a collision with the patient or the patient
chair is impossible.