7 Routine Maintenance
125
Operator Manual
Vision® 1300 Series
920515770_J
1 Chamber Manifold Connector
2 Instrument Rack Water Inlet
.
d. When rack is loaded into chamber, press Open/Close door on load side touch screen to close load door.
3. For regular DECONTAM cycle, go to Step 4.
or
If quarterly inspection shows hard water deposits are visible inside reservoir(s), proceed as follows:
IMPORTANT: Procedure to be performed by technician or biomed personnel only.
a. Put washer/disinfector Power (ON/OFF) toggle switch below touch screen display to OFF position.
Reservoirs are drained.
b. Lockout/Tagout building main electrical disconnect switch to OFF position.
c. On mechanical core inside service compartment, remove each reservoir cover.