7 — Routine Maintenance
125
Operator Manual
Vision® 1300 Series
920515770_J
1 Chamber Manifold Connector
2 Instrument Rack Water Inlet
.
d. When rack is loaded into chamber, press Open/Close door on load side touch screen to close load door.
3. For regular DECONTAM cycle, go to Step 4.
– or –
If quarterly inspection shows hard water deposits are visible inside reservoir(s), proceed as follows:
IMPORTANT: Procedure to be performed by technician or biomed personnel only.
a. Put washer/disinfector Power (ON/OFF) toggle switch below touch screen display to OFF position.
Reservoirs are drained.
b. Lockout/Tagout building main electrical disconnect switch to OFF position.
c. On mechanical core inside service compartment, remove each reservoir cover.