20
Peltier Thermostatted Single Cell Holder
326 Evolution 200 Series User Guide Thermo Scientific
Standby Mode
•Press and hold the Center button for one to two seconds to proceed to Normal Mode.
Normal Mode
Button Functions
•Press the Left button to go to the Main Menu page.
•Press the Right button to go to the Change Set Point page.
•Press and hold the Center button for one to two seconds to return to Stand-By Mode.
From the Startup page in Normal Mode, use the Change button to move about the
screen.
Y Changing the set temperature
1. Highlight Set Temperature.
The Change Set Point page is displayed with down and up arrows above the Left and
Center buttons.
Table 4. Startup modes and button functions
Page Mode Display Left button Center button Right button
Startup Standby Standby On
Normal Actual block temperature
Block Set temperature
Stirring speed (left)
Heating/cooling power (right)
Menu
(Go to Main
Menu)
Off Change (Go to
Change Set Point
Page)
Alarm Alarm description Menu Beep on/off Reset alarm/change*
*Only if the manual alarm reset option has been configured
Table 5. Normal Mode functions
Page Display Description Left button Center button Right button
Change Set
Point
Flashing selected
parameter
Change Temperature or
Stirring Speed Set Point
Down Up Change
Main Menu Parameter User Parameters Scroll Select Exit
Configure Configuration Parameters Scroll Up Exit
Diagnostic Diagnostic Values Down Up Exit

Table of Contents