25
Smart Linear 8-Cell Changer
Thermo Scientific Evolution 200 Series User Guide 355
Related Topics
Safety Considerations
Operating Precautions
Operation
Maintenance
Troubleshooting
Installing a Smart Cell Changer
It is not necessary to power off the instrument while installing or removing a Smart Accessory.
Y To install a cell changer
1. Open the sample compartment door.
2. Remove the sample compartment front panel.
3. Remove any existing cell holder or accessory.
4. Install the accessory.
The software displays a prompt to initialize the accessory.
Languages
Note The Smart Rotary Cell Changer and the Smart Linear Cell Changer cannot be used
at the same time for double-beam instruments.
CAUTION Avoid pinch hazard. Keep hands and objects clear of the accessory during
initialization.
ATTENTION Évitez tout risque de pincement. Gardez vos mains, ainsi que tout objet,
à l’écart de l’accessoire pendant l’initialisation.
VORSICHT Quetschungen sind zu vermeiden. Hände und Objekte sind während der
Initialisierung vom Zubehör fern zu halten.
CAUTELA Evitare il rischio di pizzicamento. Tenere mani e oggetti lontano
dall’accessorio durante l’inizializzazione.