12
Accutome B-Scan Plus User Guide
performed, the equipment must be turned off and the
power cord disconnected.
WARNING! THE B-SCAN PLUS IS ONLY TO
BE USED WITH A COMPUTER
USING A MEDICAL GRADE
POWER SUPPLY COMPLIANT
WITH IEC60601-1.
Please refer to the manufacturer’s instructions regarding
it’s use.
The User is responsible for ensuring
that the PC to which
the B-Scan Plus system is connected must be of the
grounding type in compliance with applicable electrical
codes.
To avoid interruption of a procedure the user is
respon
sible for ensuring that all power and USB
connections are secure and that the laptop is fully
charged.
CAUTION: To avoid possible loss of patient data
stored in the PC, ensure that the data is backed up
on some device external to the PC.
Avoiding Equipment Damage
The Accutome B-Scan Plus provides no explosion
protection from static discharge or arcing components.
Do not operate the instrument in the presence of
explosive gases such as flammable mixtures of
anesthetic and air, or nitrous oxide.