EasyManua.ls Logo

Accutome B-Scan Plus - Figure 51 Selecting Probe Frequency; Frequency; Probe Control

Accutome B-Scan Plus
268 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
69
Setup Options
Probe Control
The Probe Control options allow you to change the
default settings of the probe. The options are:
Frequency
Depth
Cine Buffer
Start/Stop
Power
Frequency
The probe frequency field allows you to select from two
frequencies: 12 MHz or 15 MHz. Which frequency you
choose for scanning will determine what depth settings
you can select. If you choose 12 MHz the depth options
are 30 mm and 60 mm. If you choose a frequency of 15
MHz the depth options are 50 mm and 100 mm. The
default probe settings are 12 MHz and 60 mm.
To change the frequency select the appropriate value
fro
m the drop down list located at the right of the
frequency field as shown in Figure 51 below.
Figure 51 Selecting Probe Frequency

Table of Contents

Related product manuals