EasyManuals Logo

Agilent Technologies 5DX Service Guide

Agilent Technologies 5DX
276 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #177 background imageLoading...
Page #177 background image
Agilent 5DX Service Guide 6-19
5DX Series 3 Pneumatic Subsystem Components
6. Using a 7/16-inch hex wrench, remove the shaft nut holding the Encoder. Refer
to Figure 6-13. Retain the shaft nut for later use.
7. Remove the Encoder.
8. Place the shaft of the new Encoder through the hole in the Encoder Bracket.
9. Using a 7/16-inch hex wrench, reinstall the nut removed in 6
10. Place the shaft of the new Encoder through the hole in the Encoder Arm
Bracket aligning the flat side of the shaft with the Encoder Set Screw, as shown
in Figure 6-14.
Figure 6-14: Positioning the Encoder Set Screw
11. Leave a 0.050-inch gap between the shaft nut and the Encoder Arm Bracket.
12. Using a 5/64-inch hex wrench, tighten the Encoder Set Screw.
13. Replace the Encoder Cover using a 1/6-inch hex wrench and the two screws
removed in 3 on page 6-18.
14. Using a 5/32-inch hex wrench, reinstall the two screws removed in 2 on page
6-17.
15. The Encoder position needs to be set. Refer to the Setting the Positions for
the Encoder procedure on page 6-20.
Set Screw
Encoder Shaft

Table of Contents

Other manuals for Agilent Technologies 5DX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5DX and is the answer not in the manual?

Agilent Technologies 5DX Specifications

General IconGeneral
BrandAgilent Technologies
Model5DX
CategoryMedical Equipment
LanguageEnglish

Related product manuals