EasyManua.ls Logo

Agilent Technologies 5DX - Page 68

Agilent Technologies 5DX
276 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-20 Safety Interlock Subsystem
Operator Control Console 5DX Series 3
9. Set the ohmmeter to the resistance mode.
10. Using a T-10 Torx driver, loosen the two screws indicated in Figure 3-14
holding the Pin Block.
11. Insert the lower side of the Primary Gap Adjuster between the Pin Block and
the Contact Block as shown in Figure 3-17. To do this, push up on the Pin
Block and insert the Primary Gap Adjuster between the Pin Block and the
Contact Block with the fingers pointing toward the Outer Barrier shielding.
Make sure that the Pin Block Bracket sits on the Primary Gap Adjuster and not
the black portion of the Pin Block
127(
The positioning of the Primary Gap Adjuster is very important. If the
Primary Gap Adjuster is positioned under the black, leading edge of
the Pin Block, then the Primary Safety Interlock will not be properly
set.

Table of Contents

Other manuals for Agilent Technologies 5DX

Related product manuals