Agilent 5DX Service Guide 3-25
5DX Series 3 Operator Control Console
9. Using a 9/16-inch open-end wrench, loosen the nut on the Switch of the
Secondary Safety Interlock as shown in Figure 3-19.
Figure 3-19: Secondary Safety Interlock
10. Attach the test leads of the ohmmeter to the test points on the Switch as shown
in Figure 3-19.
11. Set the ohmmeter to the resistance mode.
Plunger
Switch
Nut to loosen
Screws
Shroud
Test Points
T-10 Torx
Screws to
remove