11.2.1 Preparation before cleaning and disinfection
1. To avoid false activation of the wand during the cleaning and disinfection procedures, make sure to exit a scan
completely by sending the case or by going back to the home screen.
2. Remove the wand sleeve, making sure not to touch the optical surface of the wand.
Figure 202: Wand without a sleeve
3. Visually inspect the wand for any noticeable damage, for example, deterioration such as corrosion,
discoloration, pitting, or cracks.
Warning: Do not clean, disinfect, or use the wand if any damage is found. Please contact iTero Customer
Support for further instructions.
4. Prepare the following:
o
Required cleaning and disinfecting materials:
l
CaviWipes1 (or, for a list of alternative materials and the required contact time, see Approved cleaning
and disinfecting materials)
l
70% isopropyl alcohol (IPA)
l
Dry lint-free wipes
l
Soft bristle brush (e.g. the smaller end of a Healthmark Trumpet Valve Brush 1mm diameter,
Cat # 3770 or equivalent)
o
Personal Protective Equipment (PPE) and work environment
l
Please follow the cleaning and disinfection material manufacturers’ instructions
Note: Replace cleaning and disinfection materials (brushes/wipes) if visibly damaged or soiled.
Before starting the cleaning and disinfection procedure, put on your PPE.