Figure 210: iTero Element 5D Plus cart Remove remaining marks and stains on the cradle using the soft bristle brush
Figure 211: iTero Element 5D Plus mobile Remove remaining marks and stains on the cradle using the soft bristle brush
3. Using CaviWipes1, remove any remaining contaminants on the cradle.
4. Visually inspect the cradle in a well-lit area to ensure all surfaces are visibly clean.
Disinfection
l
Using CaviWipes1, thoroughly dampen all external surfaces of the cradle and ensure they remain wet for a
minimum of one (1) minute.
Note: Use multiple fresh wipes, as necessary, to keep the cradle surfaces wet for the full one (1) minute contact
time.
11.3.3 Drying cradle
Air-dry the disinfected cradle at room temperature.