Figure 210: iTero Element 5D Plus cart – Remove remaining marks and stains on the cradle using the soft bristle brush
Figure 211: iTero Element 5D Plus mobile – Remove remaining marks and stains on the cradle using the soft bristle brush
3. Using CaviWipes1, remove any remaining contaminants on the cradle.
4. Visually inspect the cradle in a well-lit area to ensure all surfaces are visibly clean.
Disinfection
l
Using CaviWipes1, thoroughly dampen all external surfaces of the cradle and ensure they remain wet for a
minimum of one (1) minute.
Note: Use multiple fresh wipes, as necessary, to keep the cradle surfaces wet for the full one (1) minute contact
time.
11.3.3 Drying – cradle
Air-dry the disinfected cradle at room temperature.
iTero Element 5D and iTero Element 5D Plus imaging systems User manual