11.6 Approved cleaning and disinfecting materials
The following table lists the Align-recommended cleaning and disinfecting materials, as well as the minimum
contact time required.
If you are using liquid disinfectant, soak a clean, sterile, lint-free wipe in the liquid and squeeze until moist, and
then follow the cleaning and disinfection instructions described in this document.
Material Active Ingredient Contact time (minutes)
CaviWipes1/CaviCide1 Quats-alcohol 1
CaviWipes/CaviCide Quats-alcohol 3
Clorox HP Wipes 1.4% Hydrogen Peroxide 5
Oxivir® 1 Wipes AHP Hydrogen Peroxide 1
Clinell Universal Range Wipes ≤50% Peracetic Acid 2
Note: If the recommended alternative disinfectants are not available in your region, consult your local supplier of
disinfecting materials for equivalent products in your region. Equivalent products must meet local regulatory
requirements, have the same active ingredients, and in addition, must be able to disinfect against at least Hepatitis
and Tuberculosis.