DRAFT
September 1, 2004 11:39 am, CH_Basic.fm
Chapter 4 Operating the Instrument
4-18 Applied Biosystems 7900HT Fast Real-Time PCR System and SDS Enterprise Database User Guide
4. When the cover has fully opened, remove the buckets from the centrifuge, then
remove the card holders from the buckets.
5. Remove all Low Density Arrays from the buckets by gently lifting them by their
carrier sides.
6. Examine the Low Density Arrays to determine whether filling is complete.
The amount of cDNA sample or control remaining in the fill reservoirs should
be uniform and consistent from reservoir to reservoir.
• If there is excess cDNA sample or control remaining in a fill reservoir (as
shown below), filling is incomplete. Proceed to the next step.
• If a fill reservoir is completely drained (as shown below), it is possible that
some wells were not filled properly.
– The Low Density Array should not be processed further. Use another
Low Density Array.
– Alternatively, you can process the Low Density Array further; however,
you should void the results for the affected fill reservoir (sample).
7. Evaluate the filling as follows:
IMPORTANT! Do not exceed 1200 rpm or accumulated centrifugation times of
more than 3 min. Excessive centrifugation speeds and times may deform the
Low Density Array.
If all fill reservoirs are... Then...
not uniform after the first
centrifuge cycle
the Low Density Array may be centrifuged again in the
same manner as before for 1 additional min.
not uniform after the
additional centrifuge cycle
• The Low Density Array should not be processed
further. Use another Low Density Array.
• Alternatively, you can process the Low Density Array
further; however, you should void the results for the
affected fill reservoir (sample).

Table of Contents

Other manuals for Applied Biosystems 7900HT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7900HT and is the answer not in the manual?

Applied Biosystems 7900HT Specifications

General IconGeneral
BrandApplied Biosystems
Model7900HT
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals