DRAFT
September 1, 2004 11:39 am, App_SWReference.fm
Operating the SDS Software from a Command Line
Applied Biosystems 7900HT Fast Real-Time PCR System and SDS Enterprise Database User Guide A-19
Enhancing the
Performance of
the SDS Software
It is possible to enhance the performance (processing speed) of the SDS software by
increasing the memory partition allocated for the software. To increase the memory
made available to the SDS software, insert the parameters shown in the example
below into any command issued to the software.
where:
• -Xms80m – Is the minimum amount of memory that the operating system will
allocate for the SDS software.
• -Xmx90m – Is the maximum amount of memory that the operating system will
allocate for the SDS software.
Applied Biosystems recommends the following settings:
• End-point runs (allelic discrimination):
-Xms20m -Xmx60m
• Real-time runs (absolute or relative quantifiation): -Xms80m -Xmx90m
jre\bin\java -Dab.home=C:\AppliedBiosystems\SDS2.2\service
-Dcom.apldbio.sds.global.approot=. -Dcom.apldbio.sds.global.config=
.\config\config.properties -Xms80m -Xmx90m -Dsource=batchfile
-Ddefport=8080…

Table of Contents

Other manuals for Applied Biosystems 7900HT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7900HT and is the answer not in the manual?

Applied Biosystems 7900HT Specifications

General IconGeneral
BrandApplied Biosystems
Model7900HT
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals