EasyManuals Logo

Applied Biosystems 7900HT User Manual

Applied Biosystems 7900HT
384 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #356 background imageLoading...
Page #356 background image
DRAFT
September 1, 2004 11:39 am, App_Theory.fm
Appendix D Theory of Operation
D-10 Applied Biosystems 7900HT Fast Real-Time PCR System and SDS Enterprise Database User Guide
Rearranging gives the following expression:
The final step is to divide the X
N
for any sample q by the X
N
for the calibrator (cb):
where:
For amplicons designed and optimized according to Applied Biosystems guidelines
(amplicon size <150 bp), the efficiency is close to 1. Therefore, the amount of target,
normalized to an endogenous reference and relative to a calibrator, is given by:
∆∆C
T
= ∆C
T,q
– ∆C
T,cb
X
N
K1E+()
C
T
∆–
×=
X
Nq,
X
Ncb,
---------------
K1( E)
C
Tq,
∆–
+×
K1( E)
C
Tcb,
∆–
+×
--------------------------------------------- -
1( E)
C
T
∆∆–
+==
2
C
T
∆∆–

Table of Contents