Operation
120
Instructions for use Primus Infinity Empowered SW 4.5n
Excess fresh gas can be discharged into the anes-
thetic gas scavenging line via the breathing system
of Primus Infinity Empowered. For this purpose,
connect the AGS hose of the non-rebreathing sys-
tem to the Y-piece of the breathing hoses con-
nected to the breathing system.
Ending the external fresh gas mode
1 Press any ventilation mode key (A).
The LED of the selected ventilation mode and
the display in the status line flash.
2 Confirm via rotary knob (B).
Ventilation via the internal rebreathing system in
Primus Infinity Empowered is restored directly in
this way.
When changing from the external non-rebreathing
system to the rebreathing system in Primus Infinity
Empowered:
z Reconnect the sample line to the Y-piece.
CAUTION
Risk of increased ambient gas concentration
Ambient air may become contaminated with an-
esthetic agent when using non-rebreathing sys-
tems.
Ensure sufficient ambient air circulation.
301
A
B
Ext. Outlet

Table of Contents

Other manuals for Dräger Primus Infinity Empowered

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Primus Infinity Empowered and is the answer not in the manual?

Dräger Primus Infinity Empowered Specifications

General IconGeneral
BrandDräger
ModelPrimus Infinity Empowered
CategoryMedical Equipment
LanguageEnglish

Related product manuals