Technical data
252
Instructions for use Primus Infinity Empowered SW 4.5n
PEEP
In Volume Mode and Volume AF
Mode
0 to 20 hPa (max. P
MAX –10 hPa)
[0 to 20 cmH
2O (max. PMAX –10 cmH2O)]
In Pressure Mode and Pressure
Support
0 to 20 hPa (max. P
INSP –5 hPa)
[0 to 20 cmH
2O (max. PINSP –5 cmH2O)]
Accuracy ±10 % of set value or ±2 hPa (±2 cmH
2O),
the greater value applies
1)
∆
PPS
In Volume Mode, Volume AF Mode
and Pressure Mode
3 to 50 hPa (max. PINSP) [3 to 50 cmH2O (max. PINSP)]
In Press. Support 3 to 50 hPa (max. P
INSP) [3 to 50 cmH2O (max. PINSP)]
0 to 2 hPa (0 to 2 cmH
2O) = Press. Support CPAP
Trigger 0.3 to 15 L per minute or OFF
T
SLOPE
In Pressure Mode, Volume AF
Mode and Pressure Support
0 to 2 seconds
Breathing system
Total volume (without breathing hoses, incl. absorber)
In Man.Spont. Typical 3.7 L
During automatic ventilation Typical 4.0 L (incl. piston volume)
Compliance (without breathing hoses)
All values apply under STPD conditions:
In Man.Spont. Typical 3.7 mL/hPa (3.7 mL/cmH
2O)
During automatic ventilation Typical 2.3 mL/hPa (2.3 mL/cmH
2O)
1) Due to gas measurement sampling and leaks (both at the patient and in the device), the end-expiratory PEEP value may
be lower than specified at the end of long expiratory phases.