Remedying faults
192
Instructions for use Primus Infinity Empowered SW 4.5n
Gas measurement failure
z Ensure adequate substitute monitoring!
Screen error
If the screen display fails:
1 Switch off Primus Infinity Empowered.
2 Set the oxygen flow to the desired value with
the O
2 emergency delivery.
3 Check vaporizer setting.
4 The patient must be ventilated manually.
5 Ensure adequate substitution monitoring.
User interface failure
If the soft keys or the rotary knob are not opera-
tional:
1 Select the Monitoring mode, see page 143.
2 The patient must be ventilated manually.