Instructions for use Primus Infinity Empowered SW 4.5n 191
Remedying faults
Ventilator and fresh gas delivery failure
The following message is displayed if both the ven-
tilator and the gas mixer fail: GAS + VENT. FAIL
(A).
The failure is indicated by a red cross through the
ventilator symbol on the screen (B) and disabled
soft keys for the ventilation mode, as well as for
fresh-gas delivery. Two prompts appear, advising
the operator how to continue:
Fresh-gas delivery failure. Check vaporizer set-
ting, open safety O
2 control valve and set a suf-
ficient flow. (C)
Ventilator failure. Manual ventilation available
only. (D)
Primus Infinity Empowered automatically switches
to the Monitoring mode.
1 Check vaporizer setting.
2 Set the safety knob of the O
2 emergency deliv-
ery to the required flow. Range: 0 to 12 L/min.
This flow streams through the vaporizer.
359
A
C
D
B
WARNING
Risk of patient injury
If the ventilator and the fresh-gas delivery fail,
the anesthesia machine switches to the venti-
lation mode Monitoring and discontinues the
fresh-gas flow automatically.
An O
2 flow must be delivered to the patient
and the patient must be ventilated manually.
Check the vaporizer setting, open the Safety
O
2 control, set a sufficient flow, set the APL
valve to an adequate pressure limiting value,
and ventilate the patient manually.
WARNING
Risk of patient injury
If pressure and volume monitoring fails, the
patient cannot be adequately monitored.
Ensure adequate substitute monitoring!