Instructions for use Primus Infinity Empowered SW 4.5n 53
Assembly and preparation
Information about transport within the clinic
Transport is defined as
– moving the device, other than for pure cali-
bration purposes.
– removing the ceiling/wall-mounted device
from the corresponding holder.
When transporting the anesthesia device:
– Only move the device using the handles
provided for this purpose.
– The anesthesia device should only be
moved by persons who are physically able.
– Dräger recommends that the anesthesia de-
vice should be moved by two persons. This
also helps to improve maneuverability.
– Take special care not to bump or knock the
device when moving it over uneven sur-
faces, around corners or at thresholds (e.g.,
in doors or elevators).
– Do not attempt to drag the device over
hoses, cables, or other obstructions on the
floor.
CAUTION
Risk of device failure
In the event of a power failure, any devices con-
nected to auxiliary power outlets will not be pow-
ered by the UPS.
Pay special attention to all power indicators of
connected devices.
CAUTION
Risk of electric shock and of device malfunction
There is a risk of injury to the user or damage to
the device if the device is connected to a power
outlet with the wrong mains voltage or without a
protective conductor.
The power cable must only be connected to a
power socket with a protective conductor, see
''Technical Data''.
007
NOTE
Risk of injury
If incorrectly handled, the anesthesia machine
may become top-heavy and tip over causing in-
jury to the patient and/or user.
Observe the following points to prevent this haz-
ard.