Instructions for use Primus Infinity Empowered SW 4.5n 63
Assembly and preparation
Connecting the patient system
WARNING
Risk of infection
Unpackaged or non-reprocessed components
might be contaminated with pathogenic
germs.
– To prevent cross-infection of patients or
users, use only new or reprocessed com-
ponents.
– Observe conditioning and assembly in-
structions.
WARNING
Risk of burns
Conductive breathing hoses or face masks
may cause burns during HF surgery.
Do not use these types of hoses and masks in
combination with HF surgery.
WARNING
Risk due to particles and dust
To protect the patient from particles and dust,
a filter must be used between the inspiratory
limb of the breathing system and the patient.
Use a Y-piece filter or filter on the inspiratory
port.
WARNING
Risk of strangulation
If not positioned with care, hoses, cables, and
similar machine components may endanger
the patient.
Take special care when connecting the pa-
tient.
CAUTION
Risk of inadequate gas concentrations
If the patient system components are not tightly
connected, ambient air will be added to the gas
mixture.
Make sure all patient system components are
tightly connected.
NOTE
Primus Infinity Empowered (without accessories)
is not made with natural rubber latex.
To minimize the risk of exposure to latex, use la-
tex-free breathing bags and breathing hoses.
NOTE
Only use original sample line – other lines may
change the technical data of the device.