Instructions for use Primus Infinity Empowered SW 4.5n
Self test results
When the self test is completed, Primus Infinity
Empowered switches to Standby.
The overall result of the self test is indicated on the
screen by a color-coded circle (A):
Dräger recommends that a full self test is carried
out before the start of therapy, after restarting the
device, and at least once every 24 hours. If the last
self test was more than 24 hours ago, the text be-
low the status display (A) is highlighted in yellow.
In addition, a message containing instructions for
further action appears in the middle of the screen
(B).
z Press the soft key self test results (C) to dis-
play more specific results.
The Self Test Results screen is displayed.
The Self Test Results screen contains the soda
lime changed (D) key.
If the soda lime is changed between cases, this key
can be pressed to log the date and time. The label
of the key then changes to undo change. The key
can be pushed again to undo the soda lime change
information. Information on when the soda lime has
been changed will be logged in the system on exit-
ing the Self Test Results screen.
This key is only available if the Infinity ID function-
ality has not been enabled by DrägerService. If the
Infinity ID functionality for the Infinity CLIC absorber
is enabled, the soda lime change can automatically
be detected by the Primus Infinity Empowered and
can be stored in the self test results.
305
Green FUNCTIONAL
Every component of the system is in
satisfactory operational order.
Yellow CONDITIONALLY FUNCTIONAL
A non-critical fault has been detected.
Primus Infinity Empowered may be
used, but call DrägerService or your lo-
cal authorized service organization.
Empty The self test was canceled.