Instructions for use Primus Infinity Empowered SW 4.5n 249
Technical data
Ambient conditions
During operation
Temperature
1)
15 to 40 °C (59 to 104 °F)
Battery charging: max. 35 °C (95 °F)
Humidity
1)
25 to 85 % (non-condensing)
Pressure
1)
700 to 1060 hPa (525 to 795 mmHg)
CO
2 concentration
1)
300 to 800 ppm
Altitude Up to 3000 m (9842 ft)
Storage/transportation
Temperature
1)
–20 to 60 °C (–4 to 140 °F)
O
2 sensor max. 50 °C (max. 122 °F)
Battery: min. –15 °C, max. 40 °C (min. 5 °F, max. 104 °F)
2)
Humidity
1)
25 to 85 % (no condensation)
Pressure
1)
500 to 1060 hPa (375 to 795 mmHg)
CO
2 concentration
1)
Not applicable
Fresh-gas delivery – electronically controlled mixer
O2 concentration 21 to 100 Vol% (for N2O as carrier gas at least 25 Vol% or
200 mL per minute – SORC)
Accuracy ±5 % or 2 Vol%; the greater value applies
Fresh-gas flow 0 and 0.2 to 18 L per minute volumetric flow
3)
Standardized to ATPD conditions.
At 0 L per minute +0.005 L per minute
Within a range of 0.20 to 0.40 L per
minute
±0.04 L per minute
Above 0.40 L per minute ±10 %
1) Depending on the type of anesthetic agent delivery unit used, this data may vary.
2) The longer-term storage at a temperature outside this range may shorten the life of the batteries.
3) The maximal permissible fresh-gas flow may be limited depending on the anesthetic agent vaporizer used.