Instructions for use Primus Infinity Empowered SW 4.5n 9
For your safety and that of your patients
Not for use in areas of explosion hazard
Connection with other electrical equip-
ment
For further information, see "Device combinations"
on page 276.
NOTE
Strictly observe the instructions for use of all ac-
cessories such as:
– Water traps
– Flow sensors
– CLIC adapter
– CLIC absorber
– Soda lime
– Breathing hoses
–Masks
– Filters
– Endotracheal suction
– Vaporizer
– Manual resuscitator
– AGSS terminal unit
WARNING
Risk of explosion and fire
The medical device is not approved for use in
areas where oxygen concentrations greater
than 25 Vol%, or combustible or explosive gas
mixtures are likely to occur.
WARNING
Risk of electric shock or device malfunction
Any connected devices or device combina-
tions not complying with the requirements
mentioned in these instructions for use can
compromise the correct functioning of the
medical device and lead to an electric shock.
– Electrical connections to equipment not
listed in these instructions for use or these
assembly instructions must only be made
when approved by each respective manu-
facturer.
– Before operating the medical device,
strictly comply with the instructions for
use of all connected devices or device
combinations.