Instructions for use Primus Infinity Empowered SW 4.5n 193
Remedying faults
System failure
If the system no longer responds to an action:
z Ventilate the patient by hand!
1 Switch Primus Infinity Empowered off and on
again: press the system power switch
(A).
2 Cancel the self test: Press the soft key cancel
test (B).
3 Select the Monitoring mode (C),
see page 143.
If the system has failed completely:
z Switch Primus Infinity Empowered off (A).
In both cases, to ensure alternative delivery of
100 % O
2 and anesthetic agent:
1 Check the vaporizer setting (D).
2 Press the safety knob (E) to unlock it and turn it
to set the flow.
042
A
B
C
D
D
E
WARNING
Risk of patient injury
If the breathing bag does not fill with fresh
gas, the patient cannot be adequately venti-
lated.
Check the oxygen supply, open cylinder
valves if necessary.
If fresh gas is still not delivered or manual ven-
tilation is not possible, close the emergency
delivery.
Disconnect the workstation from the patient
and use an alternative method of ventilation.